Agitation |
Drive |
DC motor, AC gear motor top drive motor (mechanical drive) |
Range |
50~550rpm /1rpm |
Sensor |
Magnetic hall sensor |
Impellers |
Turbine impeller (SUS 316L) Foam breaker (SUS 316L) |
Control |
Microprocessor based PID |
Temperature |
Range ( ) |
From 3 above cooling water temperature to 80 (accuracy 0.1 ) |
Control |
Temperature is automatically controlled by the external heat exchanger system |
Sensor |
RTD (Pt-100) |
Aeration |
Spurger |
Ring spurger (SUS 316L) |
Inlet filter |
Cartridge type sterilizable pleated type 0.2μmabsolute filter is applied |
Outer filter |
Same as inlet filter specification (optional item) |
Control range |
Different vessel size and/or ordered specifications |
Materials and finish |
Vessel |
Using 316L stainless steel it is polished mechanically and/or electrically |
Piping |
316L stainless steel prevents contamination and is polished by electrically |
O-rings/gaskets |
Viton or silicone material is used for O-ring and the EPDM is used for gasket |
Control unit |
Panel |
8.5" graphic Color LCD membrane keypad |
Communications |
Memory stick, printer, computer with the RS-232C communication protocol |
Controller board |
Microprocessor based integrated controller board is adopted to measure, control and display the fermentation system |
Pressure control |
Pressure is controlled manually and/or automatically |
Power supply |
220V 50/60Hz single phase, 380V 50/60Hz three phase |